Infusion Specifications

System Accuracy:
Rate Accuracy is ±5%, achieved under nominal conditions
Rate Accuracy is ±10%, achieved under low flow conditions
Occlusion Alarm Limits
Achieved under nominal conditions
Level
L2
Pressure
(mmHg)
250
approx.
Maximum Occlusion Alarm Pressure:
1038 mmHg
Post Occlusion Bolus:
Bolus volume generated at 25 ml/h when the minimum occlusion alarm threshold is reached <0.16 ml
Bolus volume generated at 25 ml/h when the maximum occlusion alarm threshold is reached <0.95 ml
Bolus Volume Accuracy:
Typical: -4.1%, Max: -3.2%, Min: -5.5% 1ml @ 10ml/h
Typical: -1.3%, Max: -0.9%, Min: -1.6% 100ml @ 1200ml/h
Maximum time for activation of occlusion alarm:
At Maximum Pressure, time to alarm at 0.1ml/h is nominally 735 [±50] minutes (Maximum <883 min)
At Minimum Pressure, time to alarm at 0.1ml/h is nominally 82 [±35] minutes (Maximum <112 min)
At Maximum Pressure, time to alarm at 1.0ml/h is nominally 65 [±4] minutes (Maximum <95 min)
At Minimum Pressure, time to alarm at 1.0ml/h is nominally 5 [+6-2] minutes (Maximum <10 min)
At Maximum Pressure, time to alarm at 25ml/h is nominally 119 [±7] seconds (Maximum <3 min)
At Minimum Pressure, time to alarm at 25ml/h is nominally 10 [+8.5-6.5] seconds (Maximum <18.5 sec)
Administering a Bolus
Parameter
Range
Bolus Rate
10 - 1200ml/h in steps of 10ml/h
Bolus Volume
0.0ml - 100.0ml in steps of 0.1ml
Displayed
Starting the Infusion / Set-up
Infusion Parameter
Range
0.1 - 99.9ml/h in steps of 0.1ml/h &
Infusion Rate
100 - 999ml/h in steps of 1ml/h
1000 - 1200ml/h in steps of 10ml/h
VTBI Primary
(0 - OFF), 1 - 9999ml
VI (Total)
0.1 - 9999ml
1a,2
1a,4
L5
L8
519
811
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
1b,2
45/50
Infusion Specifications