HP 8920A Application Handbook page 73

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

Test Set Options Required
Additional Equipment
Required
Special Test Considerations
Measurement Procedure:
1 Connect the Transmitter as shown.
CAUTION:
The RF present at the Test Set RF IN/OUT connector must not exceed 60W
continuous (or 100 Watts for 10 sec/minute).
On the Test Set:
2 Press the PRESET key.
3 Press the TX key.
Using the knob and data entry keys:
4 Set Filter 1 to 300 Hz HPF.
5 Set Filter 2 to 3 kHz LPF.
6 Select the DECODER screen.
7 Set Mode to DTMF.
8 Set Input Level to 0.95 kHz.
9 Set Gate Time to desired value (typically 100 ms).
10 Set Arm Meas to Cont.
On the Test Set:
11 Adjust Volume control to desired level.
12 Adjust Squelch control until just closed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Decoder Option
None
None
FM Transmitters
71