HP 8920A Application Handbook page 40

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

The Test Set's Features
Connectors
RF IN/OUT
DUPLEX OUT
POWER
OFF
ON
MAX POWER 60 W
!
CONTINUOUS
RF IN/OUT Connector
from the RF Generator, and input signals (60 Watts continuous, or 100
Watts for 10 sec/min) to the RF Analyzer. Nominal impedance is 50
DUPLEX OUT Connector
Generator and Tracking Generator signals. Nominal impedance is 50
ANT IN Connector
low-power RF signals (
measurements. Nominal impedance is 50
MIC/ACC Connector
connections including:
Audio microphone input for modulation of the RF output signal
Control of the RF Generator's output state
Switching between the TX TEST and RX TEST screens
Provides keying signal to control a transmitter under test
AUDIO OUT Connector
from AF Generators 1 and 2 (including encoder functions). Nominal
output impedance is%<1
AUDIO IN Connectors
signals to the AF Analyzer. Nominal impedance is 1 M
HI is the signal input for both grounding and floating input configurations.
LO may be selected to connect the signal reference to ground or float. The
connectors and controls located on the rear panel are as follows:
38
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SHIFT
CANCEL
MIC/ACC
VOLUME
ANT IN
!
MAX POWER 200 mW
type-N female connector for output signals
female BNC connector for output RF
female BNC connector for input and analysis of
200 m Watts), and for off-the-air
8-pin female DIN connector provides various
female BNC connector to output signals
at 1 kHz.
two female BNC connectors to input audio
ppm
%
Hz
ON/OFF
µV
W
dBµV
M EMORY
CARD
SQUELCH
AUDIO OUT
AUDIO IN
HI
MAX
MAX
!
!
12 v Pk
42 v Pk
.
or 600
LO
Ω.
.
.
.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8920b

Table of Contents