RF Analyzer Specifications
RF Analyzer Specifications
RF Power Measurement
Standard
Option 007
1.
To achieve the specified accuracy when measuring power at the RF IN/OUT port, the internal signal generator
level must be 60 dB below the measured power or less than -20 dBm at the DUPLEX port.
292
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1
Frequency Range: 400 kHz to 1 GHz
Measurement Range:
1 mW to 60 W continuous
100 W for 10 seconds/minute (measured at RF IN/OUT connector)
Accuracy: ±5% of reading ±0.01 mW (for temp 25° C ±10° C)
Accuracy: ±10% of reading for operating temperature range.
SWR: <1.5:1
Resolution:
Power <10 W: 1 mW
Power ≥10 W: 10 mW
Frequency Range: 400 kHz to 1 GHz
Measurement Range:
40 µW to 2.4 W continuous
4.0 W for 10 seconds/minute
Accuracy: ±5% of reading ±400 nW (for temp 25° C ±10° C)
SWR: < 1.5:1
Resolution:
P < 400 mW: 40 µW
P ≥ 400 mW: 400 µW