On the Radio set the Receiver's Controls as follows:
21. Decrease the Receiver's volume control until the AC Level meter reads
−6 dB.
On the Test Set using the knob and data entry keys:
22. Decrease Amplitude until the SINAD meter reads 12 dB.
Sensitivity (at 12dB SINAD) is displayed on the screen as Amplitude as
shown.
Sensitivity
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SINAD
0
12.57
Avg
RF Gen Freq
AFGen1 Freq
28.310700
MHz
AFGen1 To
Amplitude
0.178
FM
uV
Atten Hold
On/Off
Output Port
RF Out/Dupl
If desired, use the meter averaging function for the SINAD indicator.
a. Select dB on the SINAD meter.
b. Press the AVG key.
If desired, select the number of readings to average.
Enter the number of readings to average.
Higher number of readings averaged will require longer to