HP 8920A Application Handbook page 146

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

AM Receivers
Additional Equipment
Required
Special Test Considerations
Measurement Procedure:
1. Connect the Receiver as shown.
On the Test Set:
2. Press the PRESET key.
3. Press the RX key.
Using the knob and data entry keys:
4. Set RF Gen Freq to the receiver operating frequency.
5. Set Amplitude to −47 dBm (1 mV).
6. Set Output Port to Dupl.
7. Set AFGen1 To AM at the desired modulation depth (typically 30%).
8. Set Ext Load R to the test load resistance.
9. Set AC Level meter to measure Watts.
On the Radio set the Receiver's Controls as follows:
10. Set power to ON.
11. If required, set frequency to same value as
12. Set squelch to minimum.
13. Set RF Gain to maximum (if equipped).
14. Set the volume control until the AC Level reads 100% of the Receiver's
specified audio output.
Refer to the audio output specifications for the receiver being tested.
On the Test Set using the knob and data entry keys:
15. Select AC Level
16. Press the REF SET key.
144
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Test Load
See
"Receiver Test Loads" on page
46.