HP 8920A Application Handbook page 153

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

On the Test Set using the knob and data entry keys:
15. Select AC Level.
16. Press the REF SET key.
17. Change AFGen1 Freq from 300 Hz to 3 kHz while observing AC Level
Meter.
Frequency Response is displayed on the screen as AC Level in relative dB
(from 300Hz to 3 kHz) as shown.
The audio response should not vary more than +2 to −14 dB (Receiver's with
NOTE:
loudspeakers) or +1 to −8 dB (Receivers with headphones or feed a line) from
a standard 6 dB per octave de-emphasis curve over the frequency range of 300
to 3 kHz.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
RX TEST
AC Level
SINAD
dB
0
-0.07
24
RF Gen Freq
AFGen1 Freq
AFGen2 Freq
Filter 1
28.310000
0.5000
1.0000
50Hz HPF
MHz
kHz
kHz
AFGen1 To
AFGen2 To
Amplitude
Filter 2
-47.0
AM
FM
15kHz HPF
dBm
30.0
Off
%
Atten Hold
On/Off
Ext Load R
Output Port
RF Out/Dupl
Audio
dB
-4.72
Frequency
Ref
Response
To Screen
RF GEN
RF ANL
AF ANL
SCOPE
SPEC ANL
ENCODER
DECODER
RADIO INT
8.00
More