Using the Spectrum Analyzer
Measurement Procedure:
1. Connect the Antenna to the ANT IN connector.
CAUTION:
Do not exceed the connector's rated input or permanent instrument damage
may result.
On the Test Set:
2. Press the PRESET key.
Using the knob and data entry keys:
3. Select the SPEC ANL screen.
4. From the Controls select Ant.
5. Set Center Freq and Span fields to view desired frequency range.
6. Set Ref Level from −30 dBm to −50 dBm as required to view the desired
signal.
On the Radio:
7. Key the Transmitters and keep it keyed until the remaining steps are
completed (if applicable).