Prosense to reset password. No Field operation is available to reset password at customer
site.
Detector Model
This menu step used to set detector model. PQ series detectors named as PQD if they have display
module and PQN if they do not have display module. This menu step is used only at production and
not necessarly to change at field at any time. Once entered to menu screen will display the selected
model:
The model can be changed via up and down keys:
The detection function will not change with this setting. But detector will not report fault for
display module errors if the model is incorrect. Hence this menu step should not be altered in
Bump Test
the field.
Detectors should be tested to make sure they can perform their operations as defined. To do
this bump test should be performed in defined periods. Bump test parameters can be
adjusted via using Bump Test options in Configure menu. To start adjustments select Bump
Test step on Display menu. A new screen will display with available options:
PRS-UM-PQ-EN-Rev.05-18.06.2020
Detector Model
PQD
Detector Model
PQN
1. Test Interval
2. Inhibit Warning
3. Previous Menu
32