Installation Instructions

Refer to Figure 5-1 below for the recommended installation arrangement for the
oxygen generator.
1)
Connect the red air hose provided with the oxygen generator accessories
from the on-site air supply to the feed air inlet connection on the oxygen
generator. Refer to the Table 5-1 to determine the proper size for the feed air
inlet connection and feed air hose for your model.
Series
AS-A,AS-B
AS-D
AS-D+
AS-E
AS-G,AS-J,AS-K
AS-L
AS-N,AS-P
2)
Install your oxygen receiver. If AirSep Corporation provided your oxygen
receiver; assemble the receiver as follows (if not assembled):
AS-A—AS-P Series Instruction Manual
Red Air Hose

Figure 5.1: Typical Installation Arrangement

Feed Air Inlet Connection Size
¼" NPT female bulkhead fitting with a ¼"
NPT male x ¼" NPSM ball-end joint
adapter
¼" NPT female bulkhead fitting with a ¼"
NPT male x
" NPSM ball-end joint
adapter
3/8" NPT Female fitting
½" NPT Female fitting
¾" NPT Female fitting
1" NPT Female fitting
1-1/2" NPT Female fitting
Table 5-1 Feed Air Connection Sizes and Hose Sizes
Pressure Regulator
Kit
Oxygen Hose
Oxygen Hose
Feed Air Hose Size
¼" ID minimum
" ID minimum
3/8" ID minimum
½" ID minimum
¾" ID minimum
1" ID minimum
1-1/2" ID minimum
5-3