Installation; General - AirSep SeQual Regalia Instruction Manual

Psa oxygen concentrator system
Table of Contents

Advertisement

3.0 Installation

3.1

GENERAL

1. Select a Location
2. Inspect Inlet Air Filter
Air filter cleaning instructions are located in the maintenance section of this manual.
3. ON/OFF Button
Insert the plug into the electrical wall outlet. Push the On/Off button; the green light
will illuminate to indicate the unit is running.
4. Warm-up Period
After initially turning on the unit, allow three (3) to ten (10) minutes for the unit to
reach its specified performance. If the unit is equipped with an oxygen concentration
monitor (OM models only), the red, yellow and green light will intermittently
illuminate while the unit is warming up. For details on warm-up refer to the Display
of Operating Conditions table.
SeQual Regalia Instruction Manual
Locate oxygen concentrator in a well-ventilated space that provides
adequate airflow and check if air is being prevented from entering
the oxygen concentrator by furniture, draperies or clothing.
Always check to see if the inlet air filter is clean.
AirSep Corporation
15

Advertisement

Table of Contents
loading

Table of Contents