Final Installation - Keysight N9010AK-MTU Installation Note

Multitouch user interface upgrade kit for n9010a exa signal analyzer
Table of Contents

Advertisement

Option MTU and MTP, MultiTouch User Interface Upgrade Kits

Final Installation

Final Installation for Standard Instruments
(Benchtop Configuration,
1. Refer to
Figure
14. Carefully slide the instrument cover back onto the instrument from the rear
of the analyzer, making sure not to damage any internal cables. The seam on the cover should
be on the bottom of the instrument. Be sure the cover seats into the gasket groove in the Front
Frame Assembly.
2. Replace the four rear feet (4) to the rear of the instrument using the four screws (3). Torque
to 21 inch pounds.
3. Replace the strap handles (2) on both sides of the instrument using the four screws (1).
Torque to 21 inch pounds.
4. Locate the Top Trim Strip, Benchtop (N9020-40017) in the kit. Attach the Top Trim Strip to the
top of the Front Frame Assembly.
5. Locate the two Side Trim Strips (5041-9172) in the kit. Remove the backing and apply one
Side Trim Strip to each side of the Front Frame Assembly.
6. Replace the four instrument bottom feet.
7. Replace the four key locks to the bottom feet.
8. If the analyzer has Option EXM, External Mixer, locate the 50 Ω termination (1810-0118) that
was removed earlier. Connect the 50 Ω termination to the EXT MIXER connector.
Final Installation for Portable Instruments
(Option PRC,
Figure 15
1. Refer to
Figure
16. Carefully slide the instrument cover back onto the instrument from the rear
of the analyzer, making sure not to damage any internal cables. The seam on the cover should
be on the bottom of the instrument. Be sure the cover seats into the gasket groove in the Front
Frame Assembly.
2. Refer to
Figure
16. Replace the four rear bumpers (2) to the rear of the instrument using the
four screws (1). Torque to 21 inch pounds.
3. Refer to
Figure
15. Replace the four hole plugs (5) to both sides of the instrument.
4. Locate the Top Trim Strip, Opt PRC (N9020-40019) in the kit. Attach the Top Trim Strip to the
top of the Front Frame Assembly.
5. Using the screws (1) removed earlier, secure the four front bumpers (4) to the front frame
assembly. Torque to 21 inch pounds.
6. Refer to
Figure
15. Replace the bail handle (2) (using the four screws (1)) to the Front Frame
Assembly. Torque to 21 inch pounds.
7. If the analyzer has Option EXM, External Mixer, locate the 50 Ω termination (1810-0118) that
was removed earlier. Connect the 50 Ω termination to the EXT MIXER connector.