Assembly With Gemü Weldolet; Assembly With Gemü Grommet; Assembly With Gemü Wafer-Type Fl Ange; Assembly With Tuchenhagen Varivent Housing - GEM 3030 mFlow Operating Instructions Manual

Magnetically inductive flowmeter
Hide thumbs Also See for 3030 mFlow:
Table of Contents

Advertisement

4.1.1 Assembly with GEMÜ weldolet
Do not weld in the weld-in sleeve with the installed
flowmeter.
Lo-
cking
wing
Figure 1
Table – Installation dimensions
Figure 3
The installation depth L is calculated using the following
formula:
L = 79 - S - T
S = Wall thickness of the piping
T = Can be taken from the "Installation dimensions" table for
the respective piping
1. Drill a bolt hole with a diameter 33 in the piping.
2. Insert a weld-in sleeve (fi gure 1) into the bolt hole.
3. Align a locking wing 90° off set to the fl ow direction
(fi gure 2).
4. Weld in a weld-in sleeve with the installation dimension that
is appropriate for the nominal size (fi gure 3, see the "Instal-
lation dimensions" table) vertical to the piping axis.
5. Attach the fl owmeter (ensure that the locking wing is not
twisted).
6. Use the union nut (M33x2, SW36) to attach the fl owmeter
on the weld-in sleeve.
Figure 2
DN
T [mm]
25
4.5
32
6.0
40
7.2
50
8.3
65
10.0
80
12.0
100
14.5
125
17.5
150
20.5
200
27.0
250
33.0
300
39.5
4.1.2 Assembly with GEMÜ grommet
1. Install an in-line housing in the line.
2. Attach the fl owmeter (ensure that the locking wing is not
twisted).
3. Use the union nut (M33x2, SW36) to attach the fl owmeter
on the in-line housing.
4.1.3 Assembly with GEMÜ wafer-type fl ange
1. Install the wafer-type fl ange in the line.
2. Attach the fl owmeter (ensure that the locking wing is not
twisted).
3. Use the union nut (M33x2, SW36) to attach the fl owmeter
on the wafer-type fl ange.
4.1.4 Assembly with Tuchenhagen Varivent
housing
The electrodes of the sensor must always be
assigned precisely at 90° to the flow direction of the
medium to enable a precise measurement.
Markings
Flow
View of
bottom of
sensor
1. Weld the Varivent
in-line housing into the piping.
®
2. Position the fl owmeter with welded-on closing plate on the
housing.
3. Ensure that the sensor is aligned correctly (electrodes/
markings 90° off set to fl ow direction).
4. Use the clamp ring to secure the fl owmeter to the housing.

4.1.5 Assembly with Neumo BioControl

The electrodes of the sensor must always be
assigned precisely at 90° to the flow direction of the
medium to enable a precise measurement.
Flow
View of
bottom of
sensor
1. Weld the BioControl
®
2. Position the fl owmeter with welded-on closing plate on the
BioControl

in-line housing.

®
3. Ensure that the sensor is aligned correctly (electrodes 90°
off set to fl ow direction).
4. Screw the fl owmeter on the housing.
7 / 40

in-line

®
Closing plate
Sensor
Electrodes
in-line housing
®
Sensor
Electrodes
in-line housing into the piping.
3030 mFlow

Advertisement

Table of Contents
loading

Table of Contents