Cardinal Health Alaris PK Directions For Use Manual page 23

Syringe pump
Table of Contents

Advertisement

This option allows the 24 hour log of volume infused to be reviewed.
d
1. Press the
button to access the options menu.
2. Select the 24H LOG option using the
The display shows the hourly volume infused. The volume infused shown in brackets is the total volume infused since the volume was
last cleared. See example below:
07:48 - 08:00 4.34ml (4.34ml)
08:00 - 09:00 2.10ml (6.44ml)
09:00 - 10:00 2.10ml (8.54ml)
VOLUME CLEARED
3. Press the QUIT softkey to exit the log.
This option allows the event log to be reviewed. It can be enabled/disabled.
d
1. Press the
button to access the options menu.
2. Select the EVENT LOG option using the
3. Scroll through the log using the
To review the currently selected Data Set information:
d
1.
Press the
button to access the options menu.
2. Select DATA SET DETAILS.
3. Review the information and press the QUIT softkey to exit.
To set doserate to flowrate in precise increments, it may be necessary to switch between the rate adjust options SET BY DOSERATE and SET BY ml/
h. An arrow to the left of the rate display shows the rate changed when the
To precisely set a doserate, the arrow must be pointing to the doserate (mg/kg/h); the flowrate will be calculated from the doserate. To
precisely set a flowrate, the arrow must be pointing to the flowrate (ml/h); the doserate will be calculated from the flowrate.
Selecting the SET BY ml/h option:
1.
Whilst the pump is infusing, press the
2. Select the SET BY ml/h option using the
SET BY FLOWRATE option, the arrow on the display will automatically select the flowrate, the flowrate can be adjusted if required.
Selecting the SET BY DOSERATE option:
1.
Whilst the pump is infusing, press the
2. Select the SET BY DOSERATE option using the
SET BY DOSERATE option, the arrow on the display will automatically select the doserate, the doserate can be adjusted if required.
When in PLASMA TCI mode the user is able to switch to EFFECT SITE TCI mode if the configuration permits:
d
1.
Press the
button to access the options menu.
2. Select EFFECT SITE TCI using the
3. Press the OK softkey indicated on the screen. A confirmation screen will be displayed.
When in EFFECT SITE TCI mode the user is able to switch to PLASMA TCI mode if the configuration permits:
d
1.
Press the
button to access the options menu.
2. Select PLASMA TCI using the
3. Press the OK softkey indicated on the screen. A confirmation screen will be displayed.
Operations During Use (continued)
?
f
keys and press the OK softkey.
f
keys and press the OK softkey.
f
keys. Press the QUIT softkey to exit the log.
?
DATA SET DETAILS
?
SET BY DOSERATE/SET BY ml/h ( TIVA mode only)
d
button to access the options menu.
f
keys and press the OK softkey indicated on the screen. This will select the
d
button to access the options menu.
f
?
EFFECT SITE TCI
f
keys.
?
f
keys.
1000DF00005 Iss. 2
24 Hour Log
?
Event Log
f
keys are used to increase/decrease the infusion rate.
keys and press the OK softkey indicated on the screen. This will select the
PLASMA TCI
22/44