This option will only appear in the options menu when the infusion has been stopped.
d
1. Press the
button to access the options menu.
2. Select the END OF OPERATION option using the
3. Press the OK softkey indicated on the screen.
Note: Selecting this option will reset parameters for a new patient.
When the pump is on hold in predictive TIVA mode, the user is able to switch from TIVA to TCI mode.
d
1.
Press the
button to access the options menu.
f
keys, select the TCI MODE.
2. Using the
3. Press the OK softkey indicated on the screen. A confirmation screen will be displayed.
Note: When the mode is changed to TCI mode, the initial target will be set to zero.
When the pump is on hold in TCI mode, the user is able to switch from TCI to predictive TIVA mode.
d
1.
Press the
button to access the options menu.
f
keys, select the TIVA MODE.
2. Using the
3. Press the OK softkey indicated on the screen. A confirmation screen will be displayed.
Note: When the mode is changed to predictive TIVA mode, the initial doserate will be set to zero
In TCI and predictive TIVA mode:
d
1.
Press the
button to access the options menu.
2. Select DECREMENT CONC.
3. Select the required DECREMENT CONC and press the OK softkey to exit.
The user is able to select the Trend Size of the Concentration Prediction graph.
d
1.
Press the
button to access the options menu.
f
keys, select TREND SIZE.
2. Using the
f
keys, select the required TREND SIZE option (5 Mins, 15 Mins, 30 Mins or 60 Mins)
3. Using the
4. Press the SELECT softkey indicated on the screen.
5. Press the RESIZE softkey to rescale the vertical axis of the graph. The initial displays calculates the scale so the peak value fills graph.
If the trend is downward the graph only fills lower part and the RESIZE option forces it to rescale.
When in TCI mode, the user is able to select a numerical or graphical display.
d
1.
Press the
button to access the options menu.
f
keys, select the display mode (TEXT or GRAPH DISPLAY). The options menu shows the available display mode
2. Using the
option.
3. Press the OK softkey indicated on the screen.
d
1. Press the
button to access the options menu.
2. Select the DOSING SUMMARY option using the
3. Press the QUIT softkey to exit the menu.

Operations During Use

?
End of Operation
f
keys.
?
TCI MODE
?
TIVA MODE
?
DECREMENT CONC.
?
TREND SIZE
?
TEXT/GRAPH DISPLAY
?
Dosing Summary
f
keys and press the OK softkey.
1000DF00005 Iss. 2