Prompts - Cardinal Health Alaris PK Directions For Use Manual

Syringe pump
Table of Contents

Advertisement

Display
DOSE WOULD EXCEED
DOSE UNDER
DOSE NOT PERMITTED
TARGET WOULD EXCEED
BOLUS DOSE OVER
BOLUS DOSE UNDER
BOLUS DOSE NOT PERMITTED
WEIGHT OUTSIDE LIMIT
RATE NOT PERMITTED

Prompts

Description and Troubleshooting Guide
The infusion rate has been set to a value which exceeds a Soft Alert. Check infusion setting, to
continue with infusion at set rate press the
pressing the YES softkey. If OVERRIDE LIMIT is not required press the NO softkey and adjust
rate below Soft Alert.
The infusion rate has been set to a value which is under a Soft Alert. Check infusion setting, to
continue with infusion at set rate press the
pressing the YES softkey. If OVERRIDE LIMIT is not required press the NO softkey and adjust
rate above Soft Alert.
The infusion rate has been set above a Hard Limit. Check infusion setting and adjust rate to
appropriate required rate.
The target has been set to a value which exceeds a Soft Alert. Check infusion setting, to
continue with infusion at set target press the
by pressing the YES softkey. If OVERRIDE LIMIT is not required press the NO softkey and
adjust rate below Soft Alert.
The bolus dose has been set to a value which exceeds a Soft Alert. Check the bolus setting, to
continue with the bolus press the
the YES softkey. If OVERRIDE LIMIT is not required press the NO softkey and adjust dose
below Soft Alert.
The bolus dose has been set to a value which is under a Soft Alert. Check the bolus setting, to
continue with the bolus press the
the YES softkey. If OVERRIDE LIMIT is not required press the NO softkey and adjust dose
above Soft Alert.
The bolus dose has been set above a Hard Limit. Check bolus setting and adjust to appropriate
required dose.
The patient weight has been set to a value which exceeds or is under a Soft Alert. Check
the weight setting, to continue press the
pressing the YES softkey. If OVERRIDE LIMIT is not required press the NO softkey and adjust
the value within the limits.
The infusion rate has been set above a Hard Limit. Check infusion setting and adjust to
appropriate required rate.
1000DF00005 Iss. 2
b
button and then confi rm OVERRIDE LIMIT by
b
button and then confi rm OVERRIDE LIMIT by
b
button and then confi rm OVERRIDE LIMIT
b
button and then confi rm OVERRIDE LIMIT by pressing
b
button and then confi rm OVERRIDE LIMIT by pressing
b
button and then confi rm OVERRIDE LIMIT by
24/44

Advertisement

Table of Contents
loading