Intended use

2
Intended use
2
2
6
This  product is intended for  testing and calibration purposes on  medical devices
or systems that generate gas flows or gas pressures. That includes ventilators and
anaesthetic machines. The user of the device has received training on how to use
medical equipment and can perform repairs, maintenance and servicing on medical
devices. The device can be used in hospitals, in clinics, at device manufacturers or at
independent service companies that perform repairs or servicing on medical devices.
FlowAnalyser is intended for use in a laboratory environment. It may only be used
outside the nursing sector. It must not be used directly on patients or devices that
are connected to patients. The measuring instrument FlowAnalyser is intended for
over-the-counter sale.
FlowAnalyser is the solution for measurements in the following areas:
• Flow Low (-20 – 20 L/min)
• Flow High (-300 – 300 L/min)
• Volume
• Differential pressure
• High pressure
• Ambient pressure
• Oxygen
• Temperature
• Air humidity
• Dew point temperature
In addition, various ventilation parameters can be measured:
• Inspiratory volume, expiratory volume
• Ventilation rate
• I:E
• Inspiratory time, expiratory time
• Ppeak
• Pmean
• Pplateau
• PEEP
• PF Insp (inspiratory peak flow)
• PF Exp (expiratory peak flow)
• Ti/TCycle
• Cstat
• Delta P
FlowAnalyser is a measuring instrument for checking and calibrating
ventilators and anaesthetic machines. It must not be used for patient
monitoring. During patient treatment by the ventilator it is not allowed
to connect to FlowAnalyser.