Add Ext Mixer Connector And Cable To Front Frame Assembly - Keysight Technologies EXA N9010B Installation Notes

Signal analyzer
Hide thumbs Also See for EXA N9010B:
Table of Contents

Advertisement

Add EXT MIXER Connector and Cable to Front Frame Assembly

1. Locate the SMA female to SMA female connector (1250-1666), ¼" lock washer (2190-0067),
and ¼" hex nut (2950-0223) in the kit. If the SMA connector includes hardware, discard that
hardware and use the
2. Remove the hole-plug in the top-most hole in the upper right corner of the front frame
assembly.
3. Insert the SMA female to SMA female connector in the hole in the front frame assembly from
the front of the assembly. The hex feature on the connector should engage with the recess in
the front frame assembly. Refer to
connector using the ¼" lock washer and ¼" hex nut from the rear. Torque to 21 inch-pounds.
Figure 8
Attaching SMA Female Connector
4. Locate the External Mixing, Front Panel semi-rigid coax assembly in the kit, part number
N9020-20166. This is W27. Note that this cable is symmetrical; either end can be connected to
the front-panel connector.
5. Connect one end of W27 to the SMA female connector as shown in
to Ext Mixer
Connector,".
6. Orient W27 so that the cable slopes upward from the Ext Mixer connector at a 15 degree
angle. The top of the connector should be approximately level with the top of the shield over
the Front Panel Interface board. Refer to
cable nut on the Ext Mixer connector to 10 inch-pounds.
Installation Note N9010-90067
2190-0067
lock washer and
Figure 8, "Attaching SMA Female
Figure 10, "Proper Orientation of
Option EXM, External Mixing Upgrade Kit
2950-0223
hex nut included in this kit.
Connector,". Secure the
Figure 9, "Connecting W27
W27,". Torque the
15