12 Help messages
Code
H144
H148
H149
H150
H151
Line voltage
H152
H161
Temperature
H162
H165
H166
H170
User related
messages
H171
H172
H175
H180
H181
H182
H183
H184
H185
84
Planmeca ProMax
Explanation
Height movement is not possible
because one (or more) of the
positioning control buttons is stuck.
Height movement is not possible.
The position of the patient support
table is too high.
Height movement was stopped
because the C-arm cannot be
moved lower.
Height movement was stopped
because the patient support table
cannot be moved lower.
The line voltage was too low during
exposure.
The line voltage is too low.
The temperature of the tube head is
too high.
The temperature of the lift motor is
too high.
The temperature of the tube head is
too high for the selected exposure
values.
The maximum tube head energy
level was exceeded.
Wrong licence code.
The sensor is not attached properly
to the C-arm.
The sensor is not attached properly
to the cephalostat.
PC program selection is in conflict
with the selected X-ray unit program.
DEC is not available.
The imaging process was cancelled
in Planmeca Romexis.
Timeout in image data transmission. Exposure was interrupted.
The attached sensor is not suitable
for the selected program.
Remove the 3D sensor.
The 3D sensor is not attached
properly.
Comments
Release the button / joystick.
Press the X-ray unit down
button to move the patient
support table down.
Clear any obstruction before
continuing.
Clear any obstruction before
continuing.
Exposure was interrupted.
Contact your service
technician for help.
Exposure is not possible.
Contact your service
technician for help.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the
lift motor to cool down.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the
tube head to cool down or
use lower exposure values.
Check the licence code.
Attach and / or lock the
sensor in position.
Attach and / or lock the
sensor in position.
Select another exposure
mode in Planmeca Romexis.
Contact your service
technician for help.
Change the sensor.
Attach and / or lock the
sensor in position.