Wd1050 Dimensions - Smeg WD1050 Operating Instructions Manual

Washer disinfector
Table of Contents

Advertisement

11.

WD1050 DIMENSIONS

MACHINE DIMENSIONS [mm]
A
B
E
F
G
Legenda
"cd" – condenser drain.
"d" – drain.
"cw" – cold water (tap water).
"dw" – demi water.
NOTE: the hydraulic connections can be arranged also on the left side of the machine, always taking into
account the maximum distance specified by the product, "H".
DEMI TANK
The "demi water tank" is optional. Can be prepared if pressurized demineralized water is not
present in the plant (2bar <p_demi <5 bar). The tank must be properly positioned and used
with the accessory "PAD5".
Demineralized water helps to achieve a better washing.
IMPORTANT
If demineralized water isn't available: don't connect the demi pipe to the cold or
hot water, leave it disconnected and see the instructions (paragraph "DEMI
WATER SETTINGS" 6.11) in order to set correctly the demi water parameter.
WD1050
670
600
600
30
850
HYDRAULIC DISCHARGE
CONNECTIONS [mm]
H max
I (drain
height)
OPTIONAL DEMI TANK [mm]
L
12/04/2010
500
650 ≤ ≤ ≤ ≤ I ≤ ≤ ≤ ≤ 800
1000 ≤ L ≤ 1200
Pag. 36