Alarm ID
Description
13
Lack of demi water.
17
Water load time incorrect.
19
Demi water time exceeded.
20
Water load system fault.
22
Water flowmeter does not correctly measure.
23
Not enough water.
24
No water inside.
Low water pressure for the washing pump.
25
(Foam presence in washing chamber).
29
No drain.
30
Water safety level.
31
Water safety level fault.
WD1050
12/04/2010
Action
This alarm is active during the loading
phase.
Check if the demi water parameter is
correctly set (par. 6.11).
The alarm intervenes when the demi
turbine does not count any impulses
for 30 seconds and the state of the
pressure switch does not change.
Check the demineralized water
supply:
. cock open/closed
. tank/can empty
. solenoid valve input
. when the tank is loaded, check water
input into the tub;
. connection pipe
. connection/program correspondence
. turbine
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.)
The machine took too much time to
load the demineralized water. Check
the input pressure. For the Technical
Assistance: increase the loading
timeout (with the WD-TRACE) see
"Extra Loading Time" menu. The
extra-time has a maximum of 180
seconds.
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.).
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.)
Repeat the cycle.
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.).
Check the water supply (cock
open/closed, water pressure,
connection pipe, etc.).
Check the type of detergent used.
Repeat the cycle.
Make sure that the drain pipe is
positioned as indicated in the manual.
Repeat the cycle. Call the Technical
Assistance Service if the fault
persists.
Repeat the cycle. Call the Technical
Assistance Service if the fault
persists.