Certification - E+E Elektronik EE300Ex User Manual

Humidity / temperature sensor
Table of Contents

Advertisement

2.3 Certification

EUROPE:
The EE300Ex sensor fulfills the ATEX Directives on intrinsically safe operating equipment.
Applied standards for ATEX:
• EN 1127-1:2011
• EN 60079-0:2012
• EN 60079-11:2012
The EU-Type Examination has been carried out by TÜV SÜD Product Service GmbH.
Certified to EU-Type Examination TPS 13 ATEX 38892 003 X.
Entity parameters U
Ex-Designation
Sensor without display
Sensor with display
Remote probe
Working temperature range for the probes:
Specification of the temperature class „TKG" for use in gas area exposed to explosion hazards and
„TKD" for use in dust area exposed to explosion hazards as a function of the ambient temperature
„Tamb" for the humidity and temperature probe and the temperature probe:
Humidity and
TKG
TKD
Temperature Probe
T6
80°C
-40°C ≤ Tamb ≤ +60°C
T5
95°C
-40°C ≤ Tamb ≤ +75°C
T4
130°C -40°C ≤ Tamb ≤ +110°C
T3
195°C -40°C ≤ Tamb ≤ +175°C
T2
200°C -40°C ≤ Tamb ≤ +180°C
T1
200°C -40°C ≤ Tamb ≤ +180°C
INTERNATIONAL:
Applied Standard for IECEx:
• IEC 60079-0:2011
• IEC 60079-11:2011
The Certificate of Conformity has been carried out by FM Approvals.
Certificate No.: IECEx FMG 14.0017 X
Entity parameters:
Ex-Designation
Sensor without display
Sensor with display
Remote probe
Humidity and temperature probe:
• T6 temperature class based on -40°C (-40°F) ≤ Ta ≤ 60°C (140°F)
• T5 temperature class based on -40°C (-40°F) ≤ Ta ≤ 75°C (167°F)
• T4 temperature class based on -40°C (-40°F) ≤ Ta ≤ 110°C (230°F)
• T3 temperature class based on -40°C (-40°F) ≤ Ta ≤ 175°C (347°F)
• T2 temperature class based on -40°C (-40°F) ≤ Ta ≤ 180°C (356°F)
• T1 temperature class based on -40°C (-40°F) ≤ Ta ≤ 180°C (356°F)
8
= 28V; I
= 100mA; P
: = 700mW; C
i
i
i
I II 1G Ex ia IIC T4 Ga
I II 2G Ex ia IIC T4 Gb
II 1G Ex ia IIC T6-T1 Ga /
TKG
T6
T5
T4
T3
T2
T1
6.4 Vdc ≤ U
≤ 28Vdc; I
i
Ex ia IIC T4 Ga / Ex ia IIIC T131°C Da
Ex ia IIC T4 Gb / Ex ia IIB T4 Ga
Ex ia IIC T6-T1 Ga / Ex ia IIIC T80°C Da
= 2,2nF; L
≈ 0mH
i
i
/
II 1D Ex ia IIIC T80°C Da
/
II 1G Ex ia IIB T4 Ga
II 1D Ex ia IIIC T80°C...200°C Da
TKD
Temperature Probe
80°C
-70°C ≤ Tamb ≤ +60°C
95°C
-70°C ≤ Tamb ≤ +75°C
130°C -70°C ≤ Tamb ≤ +110°C
195°C -70°C ≤ Tamb ≤ +175°C
220°C -70°C ≤ Tamb ≤ +200°C
220°C -70°C ≤ Tamb ≤ +200°C
= 100mA; P
= 700mW; C
i
i
User Manual EE300Ex Humidity / Temperature Sensor
= 2,2nF; L
= 0mH
i
i

Advertisement