E+E Elektronik EE381 User Manual

E+E Elektronik EE381 User Manual

Moisture in oil sensor
Hide thumbs Also See for EE381:

Advertisement

Quick Links

User Manual
Moisture in Oil Sensor
EE381
BA_EE381 // v05 / Modification rights reserved // 193021

Advertisement

Table of Contents
loading

Summary of Contents for E+E Elektronik EE381

  • Page 1 User Manual Moisture in Oil Sensor EE381 BA_EE381 // v05 / Modification rights reserved // 193021...
  • Page 3: Table Of Contents

    TABLE OF CONTENTS 1. General 1.1 Explanation of Symbols 1.2 Safety Instructions 1.3 Environmental Aspects 2. Product Description 2.1 General 2.2 Dimensions 3. Installation 3.1 Installation Location 3.2 Installing the Probe directly into the Process 4. Electrical Connection 5. Operating Components 5.1 Circuit Board 5.2 Status LEDs 5.3 Display Module (Option)
  • Page 4 USA / FCC notice: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reaso- nable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the installation manual, may cause harmful interference to radio communications.
  • Page 5: General

    In addition, the manual is for everyone who requires knowledge about transport, setup, operation, maintenance and repair. ® The manual should not be used without written consent from E+E Elektronik for the purpose of competition or forwarded to third parties.
  • Page 6: Safety Instructions

    General E+E EE381 Series is specially designed for the measurement of water content in oil. EE381 is ideal for online monitoring of moisture in lubrication or insulation oil, which is very important for the long-term performance and preventive maintenance of plants and machinery.
  • Page 7: Dimensions

    (actual water content as fraction of the water content in saturated oil) = 0 corresponds to water-free oil, while a = 1 indicates saturated oil. aw measurement with the EE381 sensor is based on the outstanding long term stability and resistance to pollution of the E+E HMC capacitive sensing elements series.
  • Page 8: Installation

    User Manual EE381 Installation Installation Location Select a location that offers optimum measuring conditions. The oil must be able to circulate freely around the sensing element. Mount the sensor directly into the circulating system and not into a reservoir. Installing the Probe directly into the Process When installing a probe directly into the process, there should be a stop valve on both sides of the point of installation.
  • Page 9: Electrical Connection

    User Manual EE381 Electrical Connection Analogue output: 1 V+ 10...30V DC 2 GND 3 GND 4 OUT1 5 OUT2 6 NC 7 NC 1 V+ 10...30V DC 2 GND Page 9...
  • Page 10: Operating Components

    User Manual EE381 Operating Components Circuit Board After removal of the housing cover, the following operating components on the circuit board may be accessed to configure the sensor to the desired settings: 1. Serial interface 2. Status LEDs 3. Current / voltage output 4.
  • Page 11: Status Leds

    User Manual EE381 Status LEDs Geen (Power LED): flashing => supply voltage applied Green Yellow / Red: The yellow and red LEDs are deactivated and Yellow currently have no function. Display Module (Option) Measurand: Units: Measurand Temperature °C °F Water activity...
  • Page 12: Maintenance

    User Manual EE381 Maintenance Cleaning the Sensing Head / Filter Replacement If required, the sensing head can be cleaned. Please find the E+E Cleaning Instructions at www.epluse.com/EE381. Self-diagnosis and Error Messages Self-diagnosis via LED on the circuit board: Power LED (green): ꟷ...
  • Page 13: Replacement Parts / Accessories

    User Manual EE381 Further self-diagnosis: • Error: ꟷ possible cause  measures / help • Display shows incorrect values: • Sensor failure: ꟷ Error during re-adjustment of the sensor ꟷ No supply voltage  Reset to factory calibration and  Check wiring and supply voltage repeat the calibration routine ...
  • Page 14: Technical Data

    User Manual EE381 Technical Data Measurands Water activity Measuring range 0...1 aw Accuracy incl. hysteresis and non-linearity ±0.02 aw (0...0.9 aw) ±0.03 aw (0.9...1 aw) Temperature dependence aw: ±(0.00022 + 0.0002 x aw) x ∆T [°C] ∆T = T - 20 °C ±0.0003 °C/°C...
  • Page 15 User Manual EE381 General Supply voltage 10 - 30 V DC Current consumption, typ. voltage output 40 mA at 24 V DC current output 80 mA Pressure rating 0...20 bar (0...290 psi) 0...100 bar (0...1 450 psi) Enclosure material Aluminium die-cast (AlSi...
  • Page 16 HEADQUARTERS SUBSIDIARIES E+E Elektronik Ges.m.b.H. E+E Elektronik China E+E Elektronik Germany E+E Elektronik Korea Langwiesen 7 18F, Kaidi Financial Building, Schöne Aussicht 8 C Suite 2001, Heungdeok IT Valley 4209 Engerwitzdorf No.1088 XiangYin Road 61348 Bad Homburg Towerdong, 13, Heungdeok 1-ro,...