REMOVING THE PROBE:
• Hold the probe firmly in place. While doing so, do not
bend the probe cable.
• Release slowly closing-nut using a spanner only until
the process pressure pushes out the probe. Do not
release completely the closing-nut!
• After the probe has been pushed all the way back,
close the ball valve.
• The probe can be now removed from the ball valve.
During mounting and removal, ensure that o-ring 1 is
correctly installed.
Replace a damages o-ring by an original new one,
order code HA050308, o-ring type: 13x1.5mm
(0.5"x 0.06") - FKM-60.

3.5.4 The Optional Probe Retraction Tool

The operating instructions of the probe retraction tool must be strictly observed.
The probe of the EE300EX model T22 can be installed into a pressurized environment up to 250 bar
using the sensor retraction tools ZM-WA-025-040-EST or BG-WA-103-045-EST.
The scope of supply of the EE300EX model T22 includes the copper sealing for the Swagelok
feedthrough.
Make sure that the sensor retraction tool is in the "SERVICE" position (see manual of the sensor
retraction tool).
• Before installing the sensor, dismount the closing plug ½" using a
hexagonal 10 mm wrench and "park" it at the thread hole M10
designed for that purpose.
• Insert the probe with the copper sealing, the Swagelok
union nut, the Swagelok fitting and protective cover into
the probe retraction tool as far as it goes.
User Manual EE300Ex Humidity / Temperature Sensor
Sealing nut
Fitting body
protective cover
Swagelok-nut
Swagelok-fitting
copper sealing
Clamp sleeve
O-ring 2
O-ring 1
no aluminium
is permitted