Connect the electrode lead wires to the color-coded jacks on the side of the
preamplifier.
1
10
9
Figure 11

MB 11 Software Installation

Establishing a PC connection
To perform measurements with the MB 11 device, establishing a PC connection via USB is
required. If the MB 11 is used with office equipment that is not a medical device itself
(see section 4.4.1, PC connection 1), make sure to establish the PC connection in one
of the following ways (see section 4.4.1, PC Connection 2, 3 or 4).
WARNING
8100496 Rev. 6
Operation Manual MB 11 BERAphone
3
4
5
6
7
8
Make sure you use only office equipment with the device
that is a medical device itself or meets the requirements of
IEC 60950. If a non-medical device is used within the patient
environment (1.5 m from patient as defined in IEC 60601) a
voltage transformer must be used (exception: a battery
driven laptop is used).
1. Electrode lead wires
2. Insert Earphones IP 30
2
3. Red and blue tubes
4. Tube adapters (black for
EarCup
5. Sanibel EarCups
6. Preamplifier
7. Power / Signal quality indicator
8. Connection cable
9. MB 11 box
10. USB socket