Maico MB 11 BERAphone Operation Manual page 18

Table of Contents

Advertisement

3.3.3.3 Cleaning & Disinfection between Examinations of Different Patients
Follow this process for cleaning and disinfection of the BERAphone
1. Clean residual gel from the electrodes and ear cushion using a tissue or a
disinfectant wipe.
2. Disinfect the electrodes, ear cushion and other components that made contact
with the baby or the baby's bedding by wiping them with a fresh hospital-grade
disinfectant wipe.
a. Follow the directions for use and precautions on the disinfectant product.
b. If the disinfectant wipe is very wet, do not allow disinfectant to drip down into
the black speaker inside the BERAphone
3. Disinfect any other components that touched the baby or the baby's bedding
such as the cable, BERAphone
4. Allow the disinfectant to dry thoroughly according to the manufacturer's
instructions for maximum efficiency before using it on the next patient.
3.3.3.4 Routine Periodic Inspection and Cleaning
Figure 5 shows the design of the BERAphone
Figure 5
Daily routine periodic inspection and cleaning inside the electrode gel protectors and
beneath the ear cushion is recommended.
1. Remove the electrodes from the fixed electrode posts by pulling them straight out
(Figure 6 and Figure 7).
Figure 6
8100496 Rev. 6
Operation Manual MB 11 BERAphone
®
handle, etc.
®
.
®
.
Figure 7
16
®
/Classic
®
after each use.
23/10/2019