4 Installation and Settings

4.1 USB Specifications

The MB11 is a high-powered device that requires a USB port with
500 mA dc current over the V
- line. Passive USB hubs without an internal
USB
or separate power supply do not work with the MB 11 hardware. We
recommend using a USB port directly on the PC or notebook computer. In
most cases, these ports allow the use of high-powered functions /devices.
If you use additional USB devices on your PC and the MB11 does not work
or interferes with the function of simultaneously-used USB devices, try
another USB port to change the USB hub output.

4.2 Installation of MB 11 Software

Boot the PC and wait until the operating system finishes the start-up process.
Please note that the MB11 is only supported by MS Windows 2000 SP4 and
above and MS Windows XP SP1and above*, Windows 7 Professional and
Windows 8 Pro. The following installation guide is based on the use of MS
Windows XP. MS Vista is not supported.
Install the MB 11 Software with the USB driver files. Please note that you
need administration rights for software installation. Install the MB 11
software before connecting the MB 11 to the USB port on the computer.
After installation of the MB 11 software, connect the MB 11 to the USB port
®
of your PC and check that the red LED on the MB 11 pre-amplifier
lights
up.
If this LED does not light up, ensure that your USB port supports high-
powered USB devices or change the USB port and try again. Also see 4.1
above about using USB ports. After a few seconds, the system will show a
message about a newly found USB device "MAICO MB 11". Allow the
wizard to install the MB 11 driver if necessary.

4.3 Minimum computer specifications

Processor: Intel Pentium 800 MHz compatible or better
RAM: 256 MB
Hard disk: 2 GB
Interface: USB 1.1 or 2 (MB 11 requires a USB port with 500 mA dc current
over the VUSB – line)
Display: SVGA (800x600)
Grounded power supply
Operating system: Windows XP SP 1 or later, Windows 2000 SP 4 or later,
Windows 7 Professional and Windows 8 Pro.