Changing or Viewing
Parameters

Issuing Commands

SGD-O User Guide
This section will show you how to change a parameter using the direct
access method. In this example, change the value of the time constant. The
DAC for the time constant parameter is 007004.
1. Press: EXIT SETUP, 007004, down arrow.
2. To leave the value as is, press EXIT SETUP, and the gauge will return
to its normal display mode.
To change the value, enter the new value, and press the down arrow,
EXIT SETUP. The gauge will begin using the new value immediately
and will return to normal display mode.
This section will show you how to issue commands using the direct access
method. In this example, force the current output to 20 mA. The
command code is 5.
1. Press: EXIT SETUP, 5, down arrow.
2. The display shows two options. Press the left arrow to reject the
command or the right arrow to execute it.
Thermo Fisher Scientific