Figure 7–3. C13 in the control circuit of the FSU detector
Thermo Fisher Scientific
Times
Setting
X4
W1 jumper + W3 jumper
X5
W2 jumper + 8.25 kohm
resistor used as W3 jumper,
instead of the wire
X6
W2 jumper + 8.25 kohm
resistor used as W3 jumper
+ R36 changed to 4.12
kohm
X7
W3
X8
W2 jumper + W3 jumper
Adjusting the Gain Settings
Gain Control Settings
Comment
This parallels R5 and R37 and adds R36 to the
circuit, changing the current to the op-amp.
Using the resistor instead of the wire provides
more versatility to adjusting the current in the
feedback loop to the op-amp.
Using the resistor instead of the wire provides
more versatility to adjusting the current in the
feedback loop to the op-amp. An 8.25 kohm
resistor soldered parallel onto R36 will change
R36 to 4.12 kohm.
Everything else remains at default values.
Everything else remains at default values.
SGD-O User Guide
7-5