Softkey
f1 :
'Auto'
Used to check all groups, and to isolate the faulty board. The checking
order is shown in Figure 3- 1 3 .
f2:
Used to check the power supply group, and to isolate the faulty board in
'Power'
the power supply g roup. The items checked in the power supply group
are as follows:
f3:
' Digital'
Used to check the digital g roup (A7 , A20 to A50), and to isolate the faulty
board in the digital group. The items checked in the digital g roup are as
follows:
f4:
'Source'
Used to check the source group (A4, A5), and to isolate the faulty board i n
the source g roup. The items checked in the source g roup are a s follows:
f5:
Used to check the A6 Vector Ratio Detector (VR D) . The items checked in
'VRD'
the VRD are as follows:
f6:
'TRD'
Used to check the transducer (TRD), and isolate the faulty board in the
TRD (A2, A3). The items checked in the TRD are as fol lows:
f8:
'Exit'
Used to exit this program.
Table 3-1 4. Main Menu Softkeys
Description
Cooling fan
Output dc voltages on the A 1 board
Neon lump on the A 1 3 board
Fuses on the A 1 board
VOLTAGE SE LECTOR switch and Fuse on the rear panel
Power-on self test
Each performance test for each d igital interface
Test frequency test
Signal level/Level mon itor test
DC bias level test
H pot Buffer (A3)
A-D Converter
Prosess amp lifier and phase sensitive detector
90 0 phase shifter on A2
U N BAL detector on A2
H pot buffer on A3
MODE M/Normalizer on A2
Lcur gain amplifier on A3
Range resistor/I sense amplifier on A3
I-V converter on A3
Bias Current Isolation (Active Indactor) on A3
3-42