4-1 9. A30 HANDLER I NTERFACE (OPT.201 ) SERVICE SHEET
4- 1 9- 1 . CIRCUIT DESCRI PTIO N
The A30 handler interface board consists o f photo couplers, an d the peripheral interface
adapters.
[ Photo Couplers ]
Photo cou plers A30 U 1 through A30U7 opto-isolate the input and output signals. J umpers (
A30W1 through A30W1 1 ) are set according to the pull up voltages used. For more detail
information about the i n put/output signals and jumpers, refer to the handler interface opera­
tion note in the HP 4284A operation manual.
[ Peripheral Interface Adapters ]
A30 U 1 3 and A30 U 1 4 are peripheral i nterface adapters ( PIAs ) which interface between the
A7 CPU board and the A30 board.
4-1 9-2. TROU BLESHOOTING AIDS
The test pOints are listed i n Table 4-A30-1 .
Test Point
A30TP1
A30TP2
A30TP3
A30TP4
4-1 9-3. REPLACEABLE PARTS LISTS
The replaceable parts of the A30 handler interface board are listed in Table 4-A30-2.
4-1 9-4. COM PONENT LOCATIONS
The component locations of the A30 handler interface board and the board connector pin
assignments are shown i n Figu re 4-A30- 1 .
4-1 9-5. SCHE MATIC DIAGRAMS
The schematic d iagram of the A30 handler interface board is shown in Figure 4-A30-2.
Table 4-A30-1 . Test Point List
Signal Name
Description
Ground line
CND
5 V DC
+5 V
+
BE
792 kHz Clock signal
EXT.TRIG signal
4
A30 -1
-

Hide quick links: