Heater Test; Optic Test - RBC Bioscience MagCore Super Engineer In Training Manual

Hide thumbs Also See for MagCore Super:
Table of Contents

Advertisement

━ Heater Test

Before testing the heater, please add 500μl water
into the heat block well 1 (as shown by the arrow)
of the cartridge, and then put the cartridge into
the cartridge rack and put the thermometer in the
water.
Please select the Heater Test button.
Please select Yes to start the test.
Now the heater will be heated up to 60° C, please
check whether the temperature shown on the
thermometer is consistent with that shown on the
screen.
The acceptable temperature range is 60± 5° C.

━ Optic Test

Please put the cuvette into the cartridge rack
before the optic test without any liquid and not the
200μl SP tip.
When installing it, please do not scratch the optic test window.
Please refer to the Install MagCore® Cuvette
chapter.
Please select Optic Test.
Please select the Empty Cuvette and then wait for
15 minutes to perform the test.
The test may take some time to wait for the
D2 lamp to prewarm. The screen will show the
measured values after the test is finished. You
can insert the USB Flash Drive and press the Save
button to store the current data after the USB icon
shows in the status bar. Please take out the USB
Flash Drive and you can perform other operations
after the icon of USB Flash Drive disappears. The
acceptable range of the optical result >7000
Please save the data in the USB Flash Drive
in a personal computer, or the data in the
USB Flash Drive may be overwritten later.
File name: OpticalCuvetteCount.cvs
(MagCore Super Only)