A. Test the Tip of the W2.
1. Please put the Tip Holder Set (not provided in
accessary) into W2 of T-Rack.
2. Please add 1 ml water to the Sample Tube
and put it in W1 of the T-Rack.
3. Please select the Test W2 Tip.
4. The test can be divided into two stages.
The first stage is to observe if the piston is
smooth at taking the tips.
5. Please check whether the liquid levels of all
tips are consistent with each other.
B. Test the Tip of W3.
1. Please put the Pipette Tip into W3 of T-Rack.
2. Please add 1 ml water to the Sample Tube
and put it in W1 of the T-Rack.
3. Please select Test W3 Tip.
4. Please check whether the liquid levels are
consistent with each other. If the liquid levels
are not consistent with each other, please
check the O-Ring
chapter for more information).
C. Test the Tip of the W4.
1. Please put the 200μl SP Tip into W4 of T-Rack.
2. Please add 1 ml water to the Sample Tube
and put it in W1 of the T-Rack.
3. Please select Test W4 Tip.
4. Please check whether the liquid levels are
consistent with each other. If the liquid levels
are not consistent with each other, please
check the O-Ring
chapter for more information).
(Please refer to the Replace O-Ring
(MagCore Super Only)
(Please refer to the Replace O-Ring