━ Heater Test

Please Select Heater Test Button.
Please put the thermometer into the Heat Block
Well 1 and press Next.
Select Test Temperature
Now the heater will heat up to 65° C or 90° C,
please check whether the temperature shown on
the thermometer is consistent with what's shown
on the screen.
The acceptable temperature range is 65± 5° C or
90± 5° C.

━ UV Sterilization

Please Select the Sterilization Button.
Please select the sterilization time and press Start
button.
Warning!
The UV lamp is dangerous. Please do not
look straight at the inner of the machine
when it is operating.
Please stop operation before material
change.
Please do not open this cover by
untrained operators.
Please be cautious when UV light is on.