Replace The Timing Belt Of The Lift Motor Of The Base - RBC Bioscience MagCore Super Engineer In Training Manual

Hide thumbs Also See for MagCore Super:
Table of Contents

Advertisement

Replace the timing belt of the lift motor of the base

Open the front cover and take out the waste box, cartridge rack and T-rack.
1. Unplug the power connector to the lift motor.
2. Screw off the six screws on the liquid container above the heating module (2mm
Hex wrench).
3. Screw off the two screws on the reinforced metal sheet of the motor (3mm Hex
wrench).
4. Screw off the two screws on the bearings at the lateral grooves (4mm Hex wrench).
5. Screw off the four fixation screws on the fixation bases and move the drive module
upward (3mm Hex wrench).
6. Replace the timing belt.
7. Note that the belt should be a little tight.
8. Install the lift motor, timing belt and timing gear on the machine.
57