Verify that the handling and testing procedures were performed according to the directions in
the package insert.
-
Repeat the control test, using a new control from the same lot.
If one or more control results are still outside the specified range:
-
Perform a test using a new lot of control.
If all instructions have been followed but control results are still not within range, please
contact your Official Saladax Distributor for assistance before testing any patient
samples.

Frequency of QC Testing

Control testing is recommended when:
All quality control requirements and testing should be performed in accordance with local, state and/or
federal regulations or accreditation requirements.
QC testing is recommended when:
The Analyser is installed
A new kit shipment or lot is used
There is an unexpected patient result
New users are trained