technology means maintenance-free operation.
The Analyser is equipped with an RFID card-reader module. RFID cards are necessary for performing
all testing; they are included in the MyCare Insite Test kits and are specific to the kit. The Cuvette,
Sample Collector and RFID card are color coded. Check that the RFID card, Cuvette, and Sample
Collector are of the same color. RFID cards contain all the specific steps necessary for the Analyser to
perform tests, as well as lot data, and calibration data. The Analyser performs all the tests automatically
according to the information on the RFID card.
Test components are incubated and combined by the Analyser to generate a reaction that absorbs light
according to the amount of drug in the sample. The photometer unit performs the analysis with a light
diode. The absorption of light rays is measured during this process and the measured value is then
mathmatically converted into a test result. The result is displayed on the Touch Screen. Optionally,
results can be exported to an external computer or an HIS/LIS and can also be printed out using an
external printer.
After the test process is complete, the door opens automatically, and the test cartridge (Cuvette with
Sample Collector) can be removed and discarded. After confirming the result on the Touch Screen and
closing the Analyser door, the device is ready to perform the next test.

Manufacturer's Calibration

The MyCare Insite Analyser is manufactured according to the highest quality standards to produce safe
and accurate test results. Every Analyser is inspected and calibrated during the manufacturing
process.