52120A/Coil 6 Ka 50-Turn Coil - Fluke 5730A Calibration Manual

Multifunction
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Calibration Manual

52120A/COIL 6 KA 50-Turn Coil

Number of Turns ................................................... 50
Minimum Flexible Probe Length .......................... 500 mm
Maximum Input Current ........................................ 120 A continuous with built-in 12 V fan on
Maximum Voltage ................................................. 4.5 V rms
Input Current
0 A to 100 A
0 A to 120 A
0 A to 120 A
0 A to 120 A
0 A to 120 A
0 A to 25 A
0 A to 13 A
Notes:
1.
The inductance and mutual inductance of the 50 turn coil causes a frequency dependent compliance voltage across the coil.
Maximum frequency for 120 A input current is approximately 600 Hz. Maximum current input decreases to approximately 13 A at
10 kHz.
2.
Includes coil/probe interaction.
The specifications for these coils are at 99 % confidence level and are the combined
specification of the coil and a 52120A. If the coils are used with other current
sources the calibration specification of the coils alone is 0.65 % (99 % confidence
level) from 0 Hz to 10 kHz.
Current Output (Max.)
Input Current (Max.)
Current gain
Input Voltage (Max.)
Transconductance
Frequency
DC
<10 Hz
10 Hz to 10 kHz
Note:
The 2 A and 20 A ranges operate at full output current from DC to 10 kHz.
Frequency
DC to 850 Hz
850 Hz to 3 kHz
3 kHz to 10 kHz
28
[1]
Frequency
DC
10 Hz to 65 Hz
65 Hz to 300 Hz
300 Hz to 1 kHz
1 kHz to 3 kHz
3 kHz to 6 kHz
6 kHz to 10 kHz
2 A
2 A rms
200 mA rms
10
2 V rms
1 Siemen
120 A Range Current/Frequency Limits
Maximum Output Current
±100 A
100 A pk (70 A rms)
170 A pk (120 A rms)
Maximum Voltage Signal Applied to any Output Current Terminal with respect to Earth
600 V rms, 850 V pk, limited 2 A rms, no transient overvoltages
100 V rms, 142 V pk, limited 2 A rms, no transient overvoltages
33 V rms, 47 V pk, limited 2 A rms, no transient overvoltages
Specification
Effective Current
±(% of Amp-turns + % of 52120A range)
Amp-turns
% of Amp-turns
0 to 5000
0 to 6000
0 to 6000
0 to 6000
0 to 6000
0 to 1250
0 to 650
Note
Operating Limits
Output Current Range
20 A
20 A rms
Current Input
200 mA rms
100
Voltage Input
2 V rms
10 Siemens
Maximum Current Input
±100 mA
100 mA pk (70 mA rms)
170 mA pk (120 mA rms)
Output Isolation
52120A + Coil Specification
% of 52120A Range
0.7 %
0.7 %
0.7 %
0.7 %
0.7 %
0.7 %
0.7 %
0.7 %
0.8 %
1.0 %
1.5 %
1.0 %
5.0 %
1.0 %
120 A
120 A rms
120 mA rms
1,000
1.2 V rms
100 Siemens
Maximum Voltage Input
±1.0 V
1.0 V pk (0.7 V rms)
1.7 V pk (1.2 V rms)
[2]