Fluke 5730A Calibration Manual page 64

Multifunction
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Calibration Manual
7. Wait for the reading to settle and then record it in Table 6 under 742A STD
8. Connect the DMM to the Product as shown in Figure 7.
9. Set the Product to output the first resistance used with that standard as given
10. Push .
11. Wait for the reading to settle and then record it in the Table 6 Product Ohms
12. Set the Product to the next resistance in Table 6 that uses the same
13. Record the shown value in Table 6 under Product Displayed Value.
14. Wait for the reading to settle and then record it in Table 6.
15. Calculate the Product values using Equation 1 in Table 6 and record it under
16. Subtract the calculated value from the Product Display Value.
17. Confirm that the differences are within the limits given in the Table 6 Limits
18. Repeat the same process for the rest of the resistance measurements in
8508A Reference Multimeter
54
Ohm DMM Reading. This reading will be used with two measurements of the
Product.
in Table 6. Record the shown value in Table 6 under the Product Displayed
Value. Set the Product to 4-wire Sense On for the measurements that have
the DMM set to 4 wire and 4-wire Sense Off for the 2 wire measurements.
DMM Readings column.
resistance standard.
the Calculated Value column.
column.
Table 6.
Figure 8. 10 Ω through 19 MΩ Resistance Verification
NC
NC
742A
Resistance
Standard
HI
HI
LO
LO
hhp306.eps