3. Touch Run Wb Gain Cal to begin calibration. The Product prompts to refer to
the manual for the correct connections.
As the wideband calibration proceeds, messages are shown on the display
that identify all processes as they are encountered. When positive gains
calibration is complete, a message is shown to refer to the manual for
negative gains connections.
4. Reverse the dual-banana connector so that the center connector is
connected to LO.
5. Push . The display shows progress through the procedure in the
same fashion as for artifact calibration. At the end of the procedure, the
Product gives the option to save or discard the results.
6. Touch Save to store the results and adjust the Product. If the security
passcode has not already been entered, the Product prompts to do so before
the results are saved. Touch Discard to remove the results from the
calibration.
Calibrate the Wideband AC Module (Option 5700A-03)
Multifunction Calibrator
41