15. To reject the changes, touch Cancel. Otherwise, touch Continue to accept
16. Touch Save to store the results and adjust the Product. If the security
36
and save the changes, and to continue with calibration.
Once the internal steps are complete, a screen similar to this is shown:
Setup Menu
Calibration
Artifact Calibration
The largest change was 30% of spec.
Would you like to save changes?
Save
passcode has not already been entered, the Product prompts to do so before
the results are saved. Touch Discard to remove the results from the
calibration.
Calibration Complete
22VDC
Discard
hhp215.eps