2
Repairing the Manual Injector

Position-Sensing Switch

When required
Tools required
Parts required
1 Loosen the three stator screws.
26
No start signal when switching to the inject position
Hex key, 9/64 inch (supplied with valve)

Position-sensing switch, 0490-1849

2 Remove the stator head and stator face (see
Face"
on page 22).