5 Install the new seal. Ensure the seal is positioned as
shown.
Stator ring
7 Insert the stator face onto the stator head.
9 Secure the stator face and stator head in place with the stator screws. Tighten each screw alternately ¼-turn until the
stator head is secure.
1200 Series MI Reference Manual
6 Install the stator ring. Ensure the pin in the stator ring is
aligned with the hole in the valve body.
Rotor pin
Notch
Needle hole
8 Install the stator head and stator face onto the valve.
Ensure the pin in the stator ring is aligned with the hole in
the stator head.