Repairing the Manual Injector
7 Insert the stator face onto the stator head.
9 Secure the stator face and stator head in place with the
stator screws. Tighten each screw alternately ¼-turn until
the stator head is secure.
28
8 Install the stator head and stator face onto the valve.
Ensure the pin in the stator ring is aligned with the hole in
the stator head.
10 Reconnect the sensor cable to the start cable.