Read these instructions carefully to familiarize yourself with the product. Please retain
these operating instruction for future reference.
Use: Safety laboratory gas burners for heating and flame sterilizing. Ideal for use in
cleanroom workbenches and the laboratory.
WARNING: DO NOT LEAVE THE ACTIVATED LABORATORY
SAFETY PRECAUTIONS:
• This product is designed for exclusive use without restrictions within EU states (as
well as in other states that adhere to Directive 1999/5/EC).
• On unpacking the unit, check for possible transportation damages. Do not operate
the unit if damages are visible.
• After use or for any longer period of time without attendance, turn the main gas
supply off and turn off the gas burner at the function knob (1).
• Pay attention to your relevant rules for using liquid gas.
• Only use DVGW safety tubings with thread or tubing connectors. Check the
condition of the tube/hose frequently. Depending upon type of tube/hose, hose
clamps are required.
• All gas connections must be adequately tightened with two wrenches. Ensure gas
proofness with a suitable test fluid / equipment. DO NOT seal the swivel nut with
Teflon tape etc.
• Keep hands or other parts of the body away from the burner orifice (7a).
• Do not operate the unit near flammable liquids or hazardous materials.
• Unattended operation of the unit is not permissible.
• Always work in a well-ventilated area.
• Note that the burner orifice (7, 7a) remains hot after the flame has been
extinguished. Do not touch. Can cause burns.
• Allow sufficient time for flame orifice (7a) to cool down prior to
cleaning, desinfecting, servicing or transport. Ensure that the
unit and the gas supply are turned off.
• Because of the connectors at the back of the unit the backside should not be
sterilized with a flame.
• Allow sufficient time for burner head (7) to cool down prior to disassembling.
• Operate the unit with assembled burner head (7) only.
• After cleaning the burner head (7) allow sufficient time to dry before assembling
again.
• Keep substances away from the flame orifice (7a).
• Before mounting a nozzle check the O-Ring (20). Replace the sealing if damaged
or worn.
GAS BURNER UNATTENDED!
Need help?