6. Error messages/malfunctions

6. Error messages/malfunctions
6. Error messages/malfunctions
If an error occurs during a measurement, the
measurement is discontinued and a corresponding error
code is displayed (Example: Error No. 2).
Error No.
ERR 1
ERR 2
ERR 3
HI
LO
40
Aktualisierungsstand Gebrauchs.info November 2011
zur Verfügung gestellt von Gebrauchs.info
zur Verfügung gestellt von Gebrauchs.info
Possible cause(s)
No pulse has been detected.
Unnatural pressure impulses influence the measurement
result.
Reason: The armwas moved during the Measurement (Artefact).
The inflation of the cuff takes too long.
The cuff is not correctly seated.
The measured pulse is higher than 200 beats/min.
The measured pulse is lower than 40 beats/min.