Pc-Link Functions; Installation And Data Transmission; Error Messages/Malfunctions - Microlife BP 3AC1-1 PC Instruction Manual

Hide thumbs Also See for BP 3AC1-1 PC:
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 1

6. PC-Link Functions

This device can be used in connection with a persoanl computer (PC) running the Microlife Blood Pressure
Analyser (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable
with the PC.

6.1. Installation and Data Transmission

a)
Insert CD into CD ROM drive of your PC. The installation will
start automatically. If not, please click on «SETUP.EXE»
b)
Connect the monitor via cable with the PC – no need
to switch the device on.Three horizontal bars will
appear on the display and last for 3 seconds.
c)
The bars will then flash to indicate that the
connection between computer and device is successfully made.
As long as the cable is plugged, the bars will keep flashing and
buttons are disabled.
During the connection, the device is completly controlled by the
computer. Please refer to the «help» file for software instructions.

7. Error Messages/Malfunctions

If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(Example: Error no. 1)
Error No.
Possible cause(s)
ERR 1
The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
The tube may have become unplugged after the systolic blood pressure was measured.
Further possible cause: No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result. Reason: The arm was moved
during the measurement (Artefact).
ERR 3
Inflation of the cuff takes too long. The cuff is not correctly seated or the hose connection is
not sealed.
ERR 5
The measured readings indicated an unacceptable difference between systolic and diastolic
pressures. Take another reading following directions carefully. Contact you doctor if you
continue to get unusual readings.
ERR 6
Single data differ too much during MAM mode even after 4 cycles. No average result can be
displayed.
HI
The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200 beats
per minute). Relax for 5 minutes and repeat the measurement.*
LO
The pulse is too low (less than 40 beats per minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs repeatedly.