LoRaWAN Configuration

IMPORTANT NOTE: When you connect to the
receiver for the first time, a First-Time Setup
Wizard runs automatically. We recommend that
you close the wizard and set up the receiver
manually as described here.
Start your receiver configuration by verifying LoRaWAN
networking settings.
1. Click LoRaWAN (Network Settings).
Figure 16. LoRaWAN configuration screen
2. Ensure the following default values are entered.
a. For Smart-Vue Pro Quatro/Duo data loggers:
-
Server Address: smartvueconnect.com
-
Port (up & down): 1700
-
Keep Alive Interval: 10 seconds
CAUTION: If you enter a domain name rather
than an IP address, the DHCP client in the
receiver must be able to resolve the name to
connect to the internet (described in the next
section).
Smart-Vue Pro LoRaWAN Enabled Receiver
3. Click
in the lower right-hand corner to verify
the connection.
CAUTION: Test Lora Server will only work
once you configure your network properly (see
Setting Up an Ethernet Connection to
Setting Up a Wi-Fi Connection to your
Network).
4. When the test run has succeeded, click
your changes.
to apply