Verathon GlideScope Core Operation & Maintenance Manual page 25

Hide thumbs Also See for GlideScope Core:
Table of Contents

Advertisement

ADJUST THE MONITOR TILT ANGLE TENSION
If the angle of the monitor is difficult to adjust, or the monitor angle tilts down on its own, the workstation's
monitor angle tension needs to be adjusted.
Using a 4 mm hex wrench for a Premium Workstation, or a Phillips screwdriver for a Basic Workstation,
turn the tension adjustment screw clockwise to increase the tension, or counter‑clockwise to decrease it.
decrease
tension
ADJUST THE PREMIUM WORKSTATION ARM STRENGTH
When using the Premium Workstation, depending on the combined weight of the monitor and mounted
accessories, the strength of the articulating arm's lifting spring may need to be adjusted.
1. Lift the arm to its maximum height.
2. Using a 4 mm hex wrench, turn the spring adjustment screw clockwise to increase the arm's lifting
strength, or counter‑clockwise to decrease it. The optimal adjustment will allow the arm to remain in
position without lowering or raising on its own.
Operations & Maintenance Manual: Setting Up
Increase
tension
Weaker
Stronger
19
0900‑5017 REV‑02

Hide quick links:

Advertisement

Table of Contents