WARNING
To maintain the electrical safety of the system, the external monitors must be certified to IEC
60601‑1, IEC 60950‑1 or equivalent standards.
WARNING
To maintain electrical safety, use only the provided power supply. Connect the power cord and
power adapter to a properly grounded plug, and ensure that the disconnect is easily accessible.
Use only the accessories and peripherals recommended by Verathon.
WARNING
Electric shock hazard. Do not attempt to open the system components. This may cause serious
injury to the operator or damage to the instrument and voids the warranty. Contact Verathon
Customer Care for all servicing needs.
WARNING
Stored mechanical energy hazard. The premium workstation arm mechanism is under tension.
When the monitor is removed, it can move up rapidly on its own. Ensure that the arm is at its
highest position before removing the monitor.
WARNING
Use of accessories and cables other than those specified or provided by Verathon may cause this
system to experience electromagnetic malfunctions, including increased emissions or decreased
immunity. This may cause improper operation, procedure delays, or both.
WARNING
When equipped with a GlideScope Core 15" monitor, the GlideScope Core Premium
Workstation can tip over in some extended positions, possibly causing injury. To prevent this,
before rolling the system, move the monitor to its lowest and most retracted position.
WARNING
No modification of this equipment is allowed.
WARNING
Do not use the power adapter in the presence of flammable anesthetics.
4
0900‑5017 REV‑02